Cannabis Concentrate Sample on Raptor ARC-18 1.8 μm by UHPLC-UV
Peaks
tR (min)
1.
Δ9-Tetrahydrocannabinol (Δ9-THC)
2.49
LC_GN0592
Column
Raptor ARC-18 (cat.# 931421E)
Dimensions:
100 mm x 3.0 mm ID
Particle Size:
1.8 µm
Pore Size:
90 Å
Guard Column:
UltraShield UHPLC precolumn filter (cat.# 25809)
Temp.:
30 °C
Sample
Diluent:
25:75 Water:methanol
Inj. Vol.:
2 µL
Mobile Phase
A:
Water, 5 mM ammonium formate, 0.1% formic acid
B:
Acetonitrile, 0.1% formic acid
Time (min)
Flow (mL/min)
%A
%B
0.00
1.0
25
75
4.00
1.0
25
75
Detector
UV/Vis @ 228 nm
Instrument
UHPLC
Notes
20 mg of concentrate was weighed into a 50 mL centrifuge tube. 10 mL of ethanol was added prior to vortexing (5 seconds) and sonicating (5 minutes) over a total of three cycles. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 50-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis.
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.
I cookies consentono varie funzionalità che renderanno più agevole la tua visita al sito Restek. Utilizzando questo sito, acconsenti all’uso dei cookies in accordo con le nostre linee guida. Per saperne di più, vedi la nostra politica sulla privacy.
Restek Domestic Customer Service
Grazie.
Il messaggio è stato inviato. La ricontatteremo non appena possibile.
Messaggio non inviato.
Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037