Cannabis Hard Candy Sample on Raptor ARC-18 1.8 μm by UHPLC-UV
Peaks
tR (min)
1.
Δ9-Tetrahydrocannabinol (Δ9-THC)
2.49
LC_GN0591
Column
Raptor ARC-18 (cat.# 931421E)
Dimensions:
100 mm x 3.0 mm ID
Particle Size:
1.8 µm
Pore Size:
90 Å
Guard Column:
UltraShield UHPLC precolumn filter (cat.# 25809)
Temp.:
30 °C
Sample
Diluent:
25:75 Water:methanol
Inj. Vol.:
2 µL
Mobile Phase
A:
Water, 5 mM ammonium formate, 0.1% formic acid
B:
Acetonitrile, 0.1% formic acid
Time (min)
Flow (mL/min)
%A
%B
0.00
1.0
25
75
4.00
1.0
25
75
Detector
UV/Vis @ 228 nm
Instrument
UHPLC
Notes
1 g of ground hard candy was weighed into a 50 mL centrifuge tube. 5 mL of water was added and the sample was vortexed 30 seconds until the candy completely dissolved. Total volume was brought to 40 mL with IPA and vortexed for an additional 30 seconds. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 10-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis.
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.
I cookies consentono varie funzionalità che renderanno più agevole la tua visita al sito Restek. Utilizzando questo sito, acconsenti all’uso dei cookies in accordo con le nostre linee guida. Per saperne di più, vedi la nostra politica sulla privacy.
Restek Domestic Customer Service
Grazie.
Il messaggio è stato inviato. La ricontatteremo non appena possibile.
Messaggio non inviato.
Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037