Cannabis Hard Candy Sample on Raptor ARC-18 1.8 μm by UHPLC-UV

PeakstR (min)
1.Δ9-Tetrahydrocannabinol (Δ9-THC)2.49
image
LC_GN0591
ColumnRaptor ARC-18 (cat.# 931421E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter (cat.# 25809)
Temp.:30 °C
Sample
Diluent:25:75 Water:methanol
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.02575
4.001.02575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
Notes1 g of ground hard candy was weighed into a 50 mL centrifuge tube. 5 mL of water was added and the sample was vortexed 30 seconds until the candy completely dissolved. Total volume was brought to 40 mL with IPA and vortexed for an additional 30 seconds. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 10-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

cannabinoids

,

cannabis

,

potency

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.