Cannabis Flower Sample on Raptor ARC-18 1.8 μm by UHPLC-UV
Peaks | tR (min) | |
---|---|---|
1. | Cannabigerolic acid (CBGA)![]() | 1.19 |
Peaks | tR (min) | |
---|---|---|
2. | Δ9-Tetrahydrocannabinol (Δ9-THC)![]() | 2.49 |
3. | Tetrahydrocannabinolic acid A (THCA-A)![]() | 3.27 |

Column | Raptor ARC-18 (cat.# 931421E) | ||||||||||||
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||
Particle Size: | 1.8 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | UltraShield UHPLC precolumn filter (cat.# 25809) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Sample | |||||||||||||
Diluent: | 25:75 Water:methanol | ||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
| |||||||||||||
Detector | UV/Vis @ 228 nm | ||||||||||||
Instrument | UHPLC | ||||||||||||
Notes | 200 mg of ground flower was weighed into a 50 mL centrifuge tube. 10 mL of IPA was added prior to vortexing (5 seconds) and sonicating (5 minutes) over a total of three cycles. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 50-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis. |