16 Cannabinoids on Raptor ARC-18 1.8 μm by LC-UV

Column | Raptor ARC-18 (cat.# 931421E) | ||||||||||||
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||
Particle Size: | 1.8 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | UltraShield UHPLC precolumn filter, 0.2 μm frit (cat.# 25809) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Sample | Tetrahydrocannabivarin (cat.# 34100) | ||||||||||||
Cannabidiolic acid (CBDA) (cat.# 34099) | |||||||||||||
Cannabichromene (CBC) (cat.# 34092) | |||||||||||||
Cannabigerol (CBG) (cat.# 34091) | |||||||||||||
delta-9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093) | |||||||||||||
delta-8-Tetrahydrocannabinol ( | |||||||||||||
delta-9-Tetrahydrocannabinol ( | |||||||||||||
Cannabinol (CBN) (cat.# 34010) | |||||||||||||
Cannabidiol (CBD) (cat.# 34011) | |||||||||||||
Compounds not present in these mixes were obtained separately. | |||||||||||||
Diluent: | 25:75 Water:methanol | ||||||||||||
Conc.: | 50 µg/mL | ||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
| |||||||||||||
Detector | UV/Vis @ 228 nm | ||||||||||||
Instrument | UHPLC |