Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

LC Multiresidue Pesticide Standard #6 on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Fuberidazole2.04185.0157.0156.0
2.Ethirimol2.43210.1140.098.0
3.Flutriafol3.35302.1123.170.2
4.Nuarimol4.07315.0252.081.1
5.Paclobutrazol4.13294.1125.170.2
6.Cyproconazole isomer 14.21292.2125.170.2
7.Triadimenol4.35296.170.299.1
8.Cyproconazole isomer 24.37292.2125.170.2
9.Fluquinconazole4.44376.0348.8306.9
10.Bromuconazole isomer 14.47376.0158.970.1
11.Tetraconazole4.48372.0159.070.1
12.Triticonazole4.52318.170.1124.9
13.Etaconazole isomer 14.65328.1205.0159.0
14.Epoxiconazole4.66330.0121.0101.0
15.Fenbuconazole4.66337.0125.070.1
16.Fenarimol4.67331.0268.081.0
17.Etaconazole isomer 24.70328.1205.0159.0
18.Flusilazole4.78316.0165.0247.0
19.Bromuconazole isomer 24.89376.0158.970.1
20.Diclobutrazol4.91328.070.059.1
21.Tebuconazole4.98308.0125.070.1
22.Penconazole5.00284.070.1159.0
23.Hexaconazole5.15314.0159.070.1
24.Metconazole5.19320.1125.070.0
25.Propiconazole5.19342.0159.069.0
26.Bitertanol5.27338.1269.270.1
27.Diniconazole5.35326.170.2159.0
28.Ipconazole isomer 15.46334.2125.170.0
29.Triflumizole5.49346.0277.960.0
30.Difenoconazole5.50406.0251.1111.1
31.Ipconazole isomer 25.55334.2125.170.0
32.Etoxazole6.16360.2177.2304.2
Isomers are numbered sequentially based on elution order.
image
LC_FF0547
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:50 °C
SampleLC multiresidue pesticide standard #6 (cat.# 31977)
Diluent:Water
Conc.:20 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Methanol + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.000.40100
7.010.4955
9.500.4955
DetectorWaters Xevo TQ-S
Ion Source:Waters Zspray™ ESI
Ion Mode:ESI+
Mode:MRM
InstrumentWaters
NotesAutosampler Temperature = 5 °C
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

multiresidue

,

pesticides

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.