CARB 1004 Aldehyde/Ketone-DNPH Calibration Standard on Raptor ARC-18

PeakstR (min)
1.Formaldehyde1.516
2.Acetaldehyde1.859
3.Acetone2.283
4.Acrolein2.461
5.Propionaldehyde2.648
6.Crotonaldehyde3.177
7.Methyl ethyl ketone3.325
8.Methacrolein3.474
9.Butyraldehyde3.667
10.Benzaldehyde4.069
11.Valeraldehyde5.011
12.m-Tolualdehyde5.239
13.Hexaldehyde6.082
image
LC_EV0529
ColumnRaptor ARC-18 (cat.# 9314A1E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:40 °C
SampleCARB 1004 aldehyde/ketone-DNPH calibration standard (cat.# 33093)
Diluent:Acetonitrile
Conc.:3 µg/mL
Inj. Vol.:3 µL
Mobile Phase
A:Water:tetrahydrofuran (5:2)
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.67030
3.50.66535
6.50.63070
6.510.67030
8.50.67030
Max Pressure:240 bar
DetectorUV/Vis @ 365, 4 nm
NotesUsing stainless steel tubing and fittings instead of PEEK material is recommended when the mobile phase contains high concentrations of tetrahydrofuran.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

CARB 1004

,

aldehyde

,

ketone

,

DNPH

,

Raptor

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us