Explosives by EPA 8330B on Ultra Aromax with Trident Direct Guard
Peaks | |
---|---|
1. | HMX![]() |
2. | RDX![]() |
3. | 1,3,5-Trinitrobenzene![]() |
4. | 1,4-Dinitrobenzene (SS)![]() |
5. | 1,3-Dinitrobenzene![]() |
6. | Nitrobenzene![]() |
7. | 3,5-Dinitroaniline![]() |
8. | Tetryl![]() |
9. | 2,4,6-Trinitrotoluene![]() |
Peaks | |
---|---|
10. | Nitroglycerin![]() |
11. | 2,4-Dinitrotoluene![]() |
12. | 4-Amino-2,6-dinitrotoluene![]() |
13. | 2,6-Dinitrotoluene![]() |
14. | 2-Amino-4,6-dinitrotoluene![]() |
15. | 2-Nitrotoluene![]() |
16. | 4-Nitrotoluene![]() |
17. | 3-Nitrotoluene![]() |
18. | PETN![]() |

Column | Ultra Aromax (cat.# 9127565) |
Dimensions: | 150 mm x 4.6 mm ID |
Particle Size: | 5 µm |
Pore Size: | 100 Å |
Temp.: | 35 °C |
Sample | 8330B Nitroaromatics and nitramine mix (cat.# 33204) |
1,4-Dinitrobenzene (cat.# 33205) | |
Diluent: | Deionized water |
Conc.: | 20 µg/mL |
Inj. Vol.: | 10 µL |
Mobile Phase | Water:methanol (30:70) |
Flow: | 1.2 mL/min |
Detector | UV/Vis @ 210/254 nm |
Instrument | Shimadzu UFLCXR |
Notes | Trident direct 10 mm in-line guard cartridge holder (cat.# 25083) Ultra Aromax 10 mm x 4 mm guard cartridge (cat.# 912750210) This is the recommended confirmation column for a 2 column analysis. See LC_EV0526 for the primary column. |