Phenoxyacid Herbicides on Ultra Aqueous C18
Peaks | |
---|---|
1. | Picloram![]() |
2. | Chloramben![]() |
3. | 4-Nitrophenol![]() |
4. | Dicamba![]() |
5. | Bentazon![]() |
6. | 2,4-dichlorophenyl acetic acid (SS)*![]() |
7. | 2,4-D![]() |
8. | MCPA![]() |
9. | 3,5-Dichlorobenzoic acid![]() |
Peaks | |
---|---|
10. | MCPP![]() |
11. | Dichlorprop![]() |
12. | 2,4,5-T![]() |
13. | 2,4-DB![]() |
14. | 2,4,5-TP (Silvex)![]() |
15. | Acifluorfen![]() |
16. | 1,4-dichlorobenzene (IS)*![]() |
17. | Dinoseb![]() |
18. | Pentachlorophenol![]() |
* Not included

LC_EV0355
Column | Ultra Aqueous C18 (cat.# 9178565) | ||||||||||
Dimensions: | 150 mm x 4.6 mm ID | ||||||||||
Particle Size: | 5 µm | ||||||||||
Pore Size: | 100 Å | ||||||||||
Temp.: | ambient | ||||||||||
Sample | Chlorinated Acids by HPLC, Mix A (cat.# 32431) | ||||||||||
Chlorinated Acids by HPLC, Mix B (cat.# 32430) | |||||||||||
Diluent: | acetonitrile | ||||||||||
Conc.: | 10 ppm each analyte | ||||||||||
Inj. Vol.: | 10 µL | ||||||||||
Mobile Phase | |||||||||||
A: | 0.05% H3PO4 | ||||||||||
B: | Acetonitrile | ||||||||||
| |||||||||||
Flow: | 1.0 mL/min | ||||||||||
Detector | UV/Vis @ 225 nm | ||||||||||
Notes | Mix A = black trace Mix B = purple trace |