Cortisol and Cortisone in Extracted Female Human Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Cortisol-d40.75367.4121.0-
2.Endogenous cortisol0.76363.3120.991.1
3.Cortisone-d80.88369.4167.9-
4.Endogenous cortisone0.89361.3163.291.0
*Isobaric matrix peaks
image
LC_CF0697
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:40 °C
Sample
Diluent:Mobile phase A
Conc.: Calculated concentration is 105.1 ng/mL and 135.9 ng/mL for cortisol and cortisone, respectively
Inj. Vol.:10 µL
Mobile Phase
A:Water, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.67030
1.000.67030
1.010.60100
1.500.60100
1.510.67030
3.000.67030
Max Pressure:400 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Method:
1. Centrifuge female human urine for 5 min at 4,500 rpm, 10 °C.
2. Aliquot 380 µL supernatant. Add 20 µL each of internal standard solution (1 µg/mL in methanol).
3. Load 200 μL of sample on to ISOLUTE SLE+ 200 µL supported liquid extraction plate (part# 820-0200-P01).
4. Apply a pulse of vacuum to initiate flow.
5. Wait 5 min for sample to completely absorb.
6. Extract samples with 1 mL of MTBE. Allow solvent to flow for 5 min under gravity. Apply vacuum for 10-30 sec to complete elution.
7. Evaporate extracts to dryness under a stream of nitrogen.
8. Reconstitute in 200 µL mobile phase A prior to analysis.
9. Vortex to mix.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us