Analgesics and Non-Steroidal Anti-Inflamatory Drugs (NSAID) on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct IonESI Polarity
1.Acetaminophen1.19100152.293.065.0Positive
2.Tenoxicam2.26100338.1121.095.0Positive
3.Etoricoxib2.84100359.2280.0279.2Positive
4.Lornoxicam3.08200373.1121.195.0Positive
5.Ketorolac3.22100256.1104.977.0Positive
6.Piroxicam3.26100332.195.278.1Positive
7.Loxoprofen3.41100245.083.280.9Negative
8.Tolmetin3.64100258.1119.191.0Positive
9.Sulindac3.83100357.2233.1248.0Positive
10.Ketoprofen4.00100253.0209.0--Negative
11.Naproxen4.05200229.0168.9184.9Negative
12.Salsalate4.05100256.8136.892.9Negative
13.Meloxicam4.22100352.1115.0141.0Positive
14.Etodolac4.80100286.0241.9212.0Negative
15.Fenoprofen4.84200240.992.8--Negative
16.Oxaprozin4.86200292.0219.8102.0Negative
17.Ibuprofen4.92200205.0160.9--Negative
18.Flurbiprofen4.99100243.0199.0196.8Negative
19.Parecoxib5.20200371.2234.0165.1Positive
20.Nabumetone5.23100229.1171.1128.1Positive
21.Diclofenac5.46100293.9249.8--Negative
22.Indomethacin5.60100356.0311.8296.9Negative
23.Flufenamic acid5.75100279.9236.1176.2Negative
24.Celecoxib5.90200382.2362.0300.1Positive
25.Mefenamic acid5.91100242.1224.1--Positive
26.Tolfenamic acid6.21100259.9215.8--Negative
27.Meclofenamic acid6.29200293.9257.8177.7Negative
image
LC_CF0620
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:30 °C
Sample
Diluent:Water
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.1% formic acid + 5 mM ammonium formate
B: (95:5) Acetonitrile:water + 0.1% formic acid + 5 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.6955
1.000.67030
6.500.64060
6.510.6955
8.500.6955
DetectorMS/MS
Ion Mode:ESI+/ESI-
Mode:Scheduled MRM
InstrumentUHPLC
Notes1. Positive and negative polarity was collected simultaneously from a single injection.
2. Oxaprozin displays enhanced sensitivity using positive polarity.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us