Amphetamine and Methamphetamine (HFAA derivatized) on Rtx-5MS

Peaks
1.Amphetamine
Peaks
2.Methamphetamine
image
GC_PH00975
ColumnRtx®-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Sampleamphetamine and methamphetamine
Diluent:methanol
Conc.:100 µg/mL
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Gooseneck Splitless (4mm) w/Wool Siltek Liner
Inj. Temp.:250 °C
Oven
Oven Temp.:70 °C to 250 °C at 15 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:1 mL/min
DetectorFID @ 300 °C
NotesGooseneck Splitless (4mm) w/Wool Siltek Liner obsolete, recommend catalog number 20798

relativo alle ricerche nella libreria dei cromatogrammi

amphetamine

,

methamphetamine

,

HFAA

,

derivatized

,

heptafluorobutyric acid anhydr

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us