Steroid Sex Hormones on Rxi-1ms

Peaks
1.Androsterone
2.Dehydroepiandrosterone (DHEA)
3.17α-Estradiol
Peaks
4.Estrone
5.17β-Estradiol
6.Testosterone
7.derivatization by-product
image
GC_PH00872
ColumnRxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323)
SampleEach hormone in methanol or ethanol; compounds derivatized using 2% methoxylamine HCl (CH3ONH2) in pyridine, then N-trimethylsilylimidazole (TMSI), then analyzed
Conc.:100 µg/mL
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:3.5mm single gooseneck inlet liner
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 320 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS-Scan: Shimadzu 17A with QP5000
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:40-700 amu

relativo alle ricerche nella libreria dei cromatogrammi

hormones

,

steroids

,

estradiol

,

TMSI

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us