Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Acidic/Neutral Drugs Rxi-5ms

Peaks
1.Methprylon
2.Butalbital
3.Amobarbital
4.Meprobamate
5.Glutethimide
6.Phenobarbital
7.Methaqualone
8.Primidone
image
GC_PH00815
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
Sample1000μg/mL each
Diluent:methanol
Conc.:20 ng each compound on column
Injection
Inj. Vol.:1.0 µL split (split ratio 50:1)
Liner:Siltek Gooseneck Splitless (4mm) (cat.# 20799-214.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 220 °C at 15 °C/min to 330 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec
DetectorMS
Mode:Scan
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:35-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

acidic drugs

,

neutral drugs

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us