Dynamic Headspace USP solvents on Rtx-1301 (G43)
Peaks | tR (min) | sample concentration (µg/ml) | |
---|---|---|---|
1. | Dichloromethane![]() | 5.137 | 12.0 |
2. | Chloroform![]() | 9.260 | 1.2 |
3. | Benzene![]() | 11.145 | 0.04 |
4. | Trichloroethylene![]() | 14.601 | 1.6 |
5. | 1,4-Dioxane![]() | 17.349 | 7.6 |

Column | Rtx-1301 (G43), 30 m, 0.53 mm ID, 3.00 µm (cat.# 16085-126) |
Sample | USP <467> Calibration Mixture #5 (cat.# 36007) |
Diluent: | DMSO stock standard |
Conc.: | To each 22mL headspace vial 5ml water, ~ 1.0g of sodium sulfate and 100µL of stock standard were added. The preparations yielded the concentrations above. |
Injection | headspace-trap split |
Oven | |
Oven Temp.: | 40 °C (hold 20 min) to 240 °C at 25 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 5 mL/min |
Detector | FID @ 250 °C |
Make-up Gas Flow Rate: | 45 mL/min |
Notes | Detector: Hydrogen flow: 40 mL/min Air flow: 450 mL/min Dynamic Headspace (Trap) Conditions: Instrument: Teledyne Tekmar HT3 Valve oven temp.: 150°C Transfer line temp.: 150°C Standby flow rate: 10mL/min. Trap standby temp.: 40°C Platen/sample temp.: 80°C Sample equil. time: 15.00 min. Mixer time: 2.00 min. Mixing level: 5 Mixer stabilize time: 0.50 min. Sweep flow rate: 75mL/min. Sweep flow time: 3.00 min. Dry purge time: 5.00 min. Dry purge flow: 50mL/min. Dry purge temp.: 25°C Desorb preheat: 245°C Desorb temp.: 250°C Desorb time: 1.00 min. Trap bake temp.: 300°C Trap bake time: 6.00 min. Trap bake flow: 450mL/min. |