Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Sympathomimetic Amines (HFBA Derivatives) (Basic Drugs) on Rtx-5

Peaks
1.Amphetamine - HFBA
2.Phentermine - HFBA
3.Phenylpropanolamine - HFBA
4.Ephedrine - HFBA
5.Methamphetamine - HFBA
Peaks
6.Fenfluramine - HFBA
7.Pseudoephedrine - HFBA
8.Methylenedioxyamphetamine - HFBA
9.Methylenedioxymethamphetamine - HFBA
10.Methylphenidate - HFBA
image
GC_PH00252
ColumnRtx-5, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10223)
SampleDerivatized sympathomimetic amines
Conc.: Approximately 2.5 ng/μL
Injection
Inj. Vol.:1.0 µL splitless (hold 1.0 min)
Inj. Temp.:225 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 150 °C at 30 °C/min to 300 °C at 5 °C/min
Carrier GasHe, constant pressure
Linear Velocity:20 cm/sec @ 100 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:275 °C
Ionization Mode:EI
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Sympathomimetic

,

Amines

,

HFBA

,

Derivatives

,

Basic

,

Drugs

,

Rtx®-5

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.