Underivatized Antihistamines (Basic Drugs) on Rtx-5 Amine
Peaks | |
---|---|
1. | Phenylpropanolamine![]() |
2. | Ephedrine![]() |
3. | Pseudoephedrine |
4. | Pheniramine![]() |
5. | Diphenhydramine![]() |
6. | Doxylamine![]() |
7. | Phenyltoloxamine![]() |
8. | Methapyrilene![]() |
9. | Chlorpheniramine![]() |
10. | Brompheniramine![]() |
11. | Triprolidine![]() |

Column | Rtx-5 Amine, 30 m, 0.32 mm ID, 1.00 µm (cat.# 12354) |
Sample | |
Diluent: | Antihistamines |
Conc.: | 1000 ng/µL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Inj. Temp.: | 305 °C |
Oven | |
Oven Temp.: | 130 °C (hold 5 min) to 305 °C at 10 °C/min (hold 5 min) |
Carrier Gas | H2, constant pressure |
Linear Velocity: | 43 cm/sec @ 130 °C |
Detector | FID @ 305 °C |
Notes | FID sensitivity: 6.4 x 10-11 AFS |