Underivatized Mixed Basic Drugs on Rtx-200

Peaks
1.Nicotine
2.Benzocaine
3.Cotinine
4.Meperidine
5.Caffeine
6.Benzphetamine
7.Ketamine
8.Diphenhydramine
9.Lidocaine
10.Phenyltoloxamine
11.Tripelennamine
12.Phenothiazine
Peaks
13.Dextromethorphan
14.Methadone
15.Amitriptyline
16.Trimipramine
17.Tetracaine
18.Pyrilamine
19.Medazepam
20.Bupivacaine
21.Scopolamine
22.Codeine
23.Morphine
24.Diazepam
25.Chlorpromazine
Peaks
26.Temazepam
27.Flunitrazepam
28.Bromazepam
29.Prazepam
30.Acetopromazine
31.Flurazepam
32.Papaverine
33.Clonazepam
34.Haloperidol
35.Alprazolam
36.Triazolam
37.Thioridazine
38.Trazodone
image
GC_PH00237
ColumnRtx-200, 30 m, 0.25 mm ID, 0.25 µm (cat.# 15023)
SampleBasic drug sample
Conc.:1,000 ng/µL
Injection
Inj. Vol.:1.0 µL split (split ratio 50:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 325 °C at 4 °C/min (hold 10 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec @ 100 °C
DetectorFID @ 320 °C
NotesFID sensitivity: 1.28 x 10-10 AFS
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

underivatized

,

basic drugs

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us