Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Noble and Permanent Gases on ShinCarbon

Peaks
1.Helium
2.Neon
3.Hydrogen
4.Oxygen
5.Argon
6.Nitrogen
7.Carbon monoxide
8.Methane
9.Krypton
10.Carbon dioxide
11.Xenon
12.Sulfur hexafluoride
image
GC_PC1363
Column100/120 mesh, ShinCarbon ST micropacked column, 2 m, 1.00 mm ID (cat.# 19808)
SampleNoble and permanent gas standard
Conc.:2-10 mol%
Injection
Inj. Vol.:200 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:35 °C (hold 3 min) to 80 °C at 10 °C/min to 200 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:6.5 mL/min
DetectorTCD @ 250 °C
InstrumentAgilent 7890B GC
NotesChromatogram A Analysis Conditions:
Inj. Vol.: 50 µL, split (split ratio 10:1)
Conc.: 33 mol% each
Inj. Temp.: 150 °C
Oven Temp.: 35 °C isothermal
Carrier Gas: N2, constant flow
Flow Rate: 5.5 mL/min
Detector: TCD @ 250 °C
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us