Jet Fuel on Rxi-1ms by GC-VUV (ASTM D8267)

GC_PC1359
Column | Rxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323) |
Sample | Jet fuel |
Conc.: | Neat |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 50 °C (hold 0.1 min) to 260 °C at 15 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 2 mL/min |
Detector | VUV |
Transfer Line Temp.: | 275 °C |
Flow Cell Temp.: | 275 °C |
Acquisition Range: | 125-240 nm |
Acquisition Rate: | 7 spectra/sec |
Instrument | HP 6890 GC & VUV Analytics VGA-100 |
Notes | ASTM D8267, “Standard Test Method for Determination of Saturated Hydrocarbon, Aromatic, and Diaromatic Content of Aviation Turbine Fuels Using Gas Chromatography with Vacuum Ultraviolet Absorption Spectroscopy Detection (GC-VUV),” is used to determine total concentration of saturated hydrocarbon, aromatic, and diaromatic components of aviation turbine engine fuels (jet fuel), not individual compounds. |
Acknowledgement | Restek would like to thank Jack Cochran from VUV Analytics for sharing the data. |