Thiophene in Benzene on Rtx-Wax

Peaks
1.Elution of benzene
2.Thiophene
image
GC_PC1357
ColumnRtx-Wax, 30 m, 0.32 mm ID, 1.00 µm (cat.# 12454)
Sample
Diluent:Benzene
Conc.:0.5 mg/kg
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 100 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorSCD
InstrumentAgilent 7890B GC
NotesBenzene is a popular industrial solvent and a precursor to many other aromatic chemicals. Determination of thiophene in benzene by ASTM D7011 is achieved using a sulfur selective detector. Most sulfur selective detectors will show signs of quenching at the point when the hydrocarbon solvent elutes. Therefore, for successful analysis, it is vital to separate thiophene from benzene. Rtx-Wax columns provide the required inertness and proper selectivity needed to separate those two compounds.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

ASTM D7011

,

purity analysis

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us