Permanent Gases on Rt-Msieve 5A (15 m x 0.25 mm x 20 μm)

PeakstR (min)Conc. (Mol%)
1.Helium0.58Balance
2.Argon0.883
3.Oxygen0.912
4.Nitrogen1.745
5.Methane2.344
6.Carbon monoxide7.625
image
GC_PC1349
ColumnRt-Msieve 5A, 15 m, 0.25 mm ID, 20 µm (cat.# 19773)
SamplePermanent gases (custom mix)
Injection
Inj. Vol.:30 µL split (split ratio 40:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:30 °C (hold 10 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
DetectorTCD @ 200 °C
InstrumentAgilent 7890B GC
NotesResolution (Ar/O2) = 1.7
Asymmetry (CO) = 1.2

Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

MolSieve

,

Molecular Sieve

,

Natural Gas

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us