Permanent Gases and Nitrous Oxide on ShinCarbon ST Micropacked Column

PeakstR (min)
1.Air (oxygen, nitrogen)0.48
2.Carbon monoxide0.54
3.Methane0.83
4.Carbon dioxide1.30
PeakstR (min)
5.Nitrous oxide1.65
6.Ethyne (acetylene)2.48
7.Ethene (ethylene)3.37
8.Ethane4.51
image
GC_PC1343
ColumnShinCarbon ST micropacked column, 100/120 mesh, 2 m, 1.00 mm ID (cat.# 19808)
Sample
Conc.: Varied (0.1-0.5% by volume)
Injection
Inj. Vol.:100 µL split (split ratio 5:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:200 °C
Oven
Oven Temp.:165 °C (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:15 mL/min
DetectorTCD @ 200 °C
InstrumentAgilent 7890B GC
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

greenhouse gasses

,

permanent gases

,

environmental pollutant

,

drug of abuse

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.