Permanent Gases on Capillary Rt-MSieve 5A
Peaks | tR (min) | Conc. mol % | |
---|---|---|---|
1. | Hydrogen![]() | 1.358 | 4 |
2. | Argon![]() | 1.976 | 3 |
3. | Oxygen![]() | 2.028 | 2 |
4. | Nitrogen![]() | 3.528 | 5 |
5. | Methane![]() | 4.755 | 4 |
6. | Carbon monoxide![]() | 12.499 | 5 |

Column | Rt-Msieve 5A, 30 m, 0.32 mm ID, 30 µm (cat.# 19722) |
Sample | Permanent gas standard |
Diluent: | Helium |
Injection | |
Inj. Vol.: | 50 µL split (split ratio 20:1) |
Liner: | Topaz 2.0 mm ID straight inlet liner (cat.# 23313) |
Inj. Temp.: | 150 °C |
Oven | |
Oven Temp.: | 40 °C (hold 13 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.5 mL/min |
Detector | TCD @ 200 °C |
Make-up Gas Flow Rate: | 10 mL/min |
Make-up Gas Type: | He |
Instrument | Agilent 7890B GC |
Notes | Baseline resolution between argon and oxygen is obtained using split injection. (R(Ar/O2) = 1.8) Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution. |