Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Permanent Gases on Capillary Rt-MSieve 5A

PeakstR (min)Conc. mol %
1.Hydrogen1.3584
2.Argon1.9763
3.Oxygen2.0282
4.Nitrogen3.5285
5.Methane4.7554
6.Carbon monoxide12.4995
image
GC_PC1331
ColumnRt-Msieve 5A, 30 m, 0.32 mm ID, 30 µm (cat.# 19722)
SamplePermanent gas standard
Diluent:Helium
Injection
Inj. Vol.:50 µL split (split ratio 20:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:40 °C (hold 13 min)
Carrier GasHe, constant flow
Flow Rate:2.5 mL/min
DetectorTCD @ 200 °C
Make-up Gas Flow Rate:10 mL/min
Make-up Gas Type:He
InstrumentAgilent 7890B GC
NotesBaseline resolution between argon and oxygen is obtained using split injection. (R(Ar/O2) = 1.8)

Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

MolSieve

,

Natural Gas

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.