Sulfur Compounds on Rtx-1

Column | Rtx-1, 30 m, 0.32 mm ID, 4.00 µm (cat.# 10198) |
Sample | ASTM D5623 standards 1 & 2 from DCG Partnership 1, Ltd. |
Injection | |
Inj. Vol.: | 0.2 µL split (split ratio 10:1) |
Liner: | 4 mm ID cup inlet liner with Siltek deactivation |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 24 mL/min |
Oven | |
Oven Temp.: | 30 °C (hold 2 min) to 200 °C at 10 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.4 mL/min |
Linear Velocity: | 38 cm/sec @ 30 °C |
Detector | DP SCD @ 800 °C |
Burner Vacuum: | 363 torr |
Hydrogen Flow Rate: | 40 mL/min |
Air Flow Rate: | 60 mL/min |
Instrument | Agilent 7890B GC |
Notes | Solvent standard of sulfur components found in gasoline samples utilizing ASTM D5623. |