Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Stability Overlay: 1st and 40th Injections of Short-Chain Amines in Water on Rtx-Volatile Amine

Peaks
1.Methanol
2.Dimethylamine
3.Trimethylamine
4.Methylethylamine
5.Dimethylethylamine
6.Diethylamine
7.Methyldiethylamine
8.Triethylamine
image
GC_PC1244
ColumnRtx-Volatile Amine, 60 m, 0.32 mm ID (cat.# 18078)
SampleShort-chain amines
Diluent:Water
Injection
Inj. Vol.:1 µL split (split ratio 15:1)
Inj. Temp.:220 °C
Split Vent Flow Rate:30 mL/min
Oven
Oven Temp.:40 °C (hold 10 min) to 250 °C at 20 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
Linear Velocity:35 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
AcknowledgementCourtesy of Gilbert Baele, Taminco (Antwerp, Belgium)
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us