Short-Chain Amines on Rtx-Volatile Amine

Peaks
1.Monomethylamine (MMA)
2.Methanol
3.Dimethylamine (DMA)
Peaks
4.Ethylamine
5.Trimethylamine (TMA)
6.Dimethylethylamine
7.Methyldiethylamine
image
GC_PC1243
ColumnRtx-Volatile Amine, 60 m, 0.32 mm ID (cat.# 18078)
SampleShort-chain amines
Diluent:Water
Injection
Inj. Vol.:1 µL split (split ratio 15:1)
Inj. Temp.:220 °C
Split Vent Flow Rate:30 mL/min
Oven
Oven Temp.:40 °C (hold 10 min) to 250 °C at 20 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
Linear Velocity:35 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
AcknowledgementCourtesy of Gilbert Baele, Taminco (Antwerp, Belgium)