Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Sulfinert Treatment Prevents Loss of Sulfur Compounds

PeaksConcentration (ppbv)
1.Hydrogen sulfide20
2.Methyl mercaptan200
3.Ethyl mercaptan300
4.Dimethyl sulfide2,000
5.2-Propanethiol50
6.1-Propanethiol100
7.Ethyl methyl sulfide1,300
8.2-Butanethiol150
9.Diethyl sulfide180
image
GC_PC1176
ColumnRtx-1, 60 m, 0.53 mm ID, 7.00 µm (cat.# 10193)
SampleNatural gas
InjectionDirect
Inj. Temp.:30 °C
Oven
Oven Temp.:30 °C (hold 1 min) to 100 °C at 30 °C/min (hold 60 min)
Carrier GasHe, constant pressure
Linear Velocity:96 cm/sec @ 30 °C
DetectorSievers 355 SCD @ 800 °C
Hydrogen Flow Rate:100 mL/min
Air Flow Rate:40 mL/min
InstrumentHP5890 GC
NotesPoint of use samples were collected in either an uncoated sampling cylinder or a Sulfinert-treated sampling cylinder. A Valco sampling valve with a 1 mL external loop was used.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

sulfinert

,

sample cylinder

,

cylinder

,

sulfur

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.