DHA Column Inertness Comparison Using Oxy Setup

Peaks
1.Ethanol
2.Pentane (C5)
3.tert-Butanol
4.2-Methylbutene-2
image
GC_PC01098
ColumnComparison of competitor column and Rtx-DHA-100 column, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleOxy setup blend, neat
Injection
Inj. Vol.:0.1 µL split (split ratio 150:1)
Liner:4 mm ID deactivated cup inlet liner (cat.# 20709 [replaced by 20710])
Inj. Temp.:250 °C
Oven
Oven Temp.:5 °C
Carrier GasH2, constant flow
Flow Rate:3.62 mL/min
Linear Velocity:55 cm/sec
DetectorFID @ 300 °C
NotesThe chromatograms are aligned on pentane (C5).

Note that the analytical column was originally named Rtx-1 PONA (cat.# 10195); it has since been renamed (no other changes) Rtx-DHA-100 (cat.# 10148).
Download PDF