Gasoline Containing Ethanol by ASTM D3606-06 (Modified) on D3606 Application Column Set
Peaks | |
---|---|
1. | Ethanol![]() |
2. | Benzene![]() |
3. | 2-Butanol (IS)![]() |
4. | Toluene![]() |

Column | D3606 Application Column (2 column set) Column 1: 6' (1.8m), 1/8" OD, 2.0mm ID, nonpolar Rtx®-1 Column 2: 16' (4.9m), 1/8" OD, 2.0mm ID, proprietary packing material (cat.# 83606-800) |
Sample | Gasoline with internal standard |
Diluent: | |
Injection | sample valve |
Sample Loop Vol.: | 1.5 µL |
Oven | |
Oven Temp.: | 135 °C (hold 16 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 20.0 mL/min |
Detector | TCD @ 200 °C |
Notes | 3 min. backflush |
Acknowledgement | Chromatogram courtesy of Boguslaw Dudek, Conoco Phillips, Linden, NJ |