Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Derivatized B100 and Internal Standards on MXT-Biodiesel TG with 2m Integra Gap

image
GC_PC00934
ColumnMXT-Biodiesel TG 13m, 0.53mm ID, 0.16µm with built-in 2m Integra-Gap (total column length 15 m)
SampleB100 + IS Butanetriol & Tricaprin derivatized with MSTFA as per ASTM D-6584
Injection
Inj. Vol.:1.0 µL cold on-column
Temp. Program:oven track
Oven
Oven Temp.:50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 430 °C at 30 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:4 mL/min
DetectorFID @ 430 °C
InstrumentShimadzu 2010 GC
Notes(Data acquired on prototype column, similar column to cat# 70290)

relativo alle ricerche nella libreria dei cromatogrammi

glycerides

,

total glycerides

,

biodiesel

,

B100

,

ASTM D-6584

,

ASTM

,

6584

,

triglycerides

,

EN14105

,

residual triglyceride

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us