Biodiesel Oil (B100) and Internal Standards on Rtx-Biodiesel TG

image
GC_PC00908
ColumnRtx-Biodiesel TG with Hydroguard guard column (2m x 0.53mm id) connected with Alumaseal connector, 10 m, 0.32 mm ID, 0.10 µm (cat.# 10291)
Samplebiodiesel oil (B100) plus internal standards butanetriol and tricaprin, and derivatized per ASTM Method D 6584
Injection
Inj. Vol.:1 µL cold on-column
Temp. Program:oven track
Oven
Oven Temp.:50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:4 mL/min
DetectorFID @ 380 °C
InstrumentAgilent/HP6890 GC
Download PDF