Fast Detailed Hydrocarbons Analysis (DHA) on Rtx-DHA-100/Rtx-5 DHA

Peaks
1.Ethanol
2.Pentane (C5)
3.tert-Butanol
4.2-Methylbutene-2
5.2,3-Dimethylbutane
6.Methyl tert-butyl ether (MTBE)
7.Hexane (C6)
8.1-Methylcyclopentene
9.Benzene
10.Cyclohexane
11.3-Ethylpentane
12.1,2-Dimethylcyclopentane
13.Heptane (C7)
14.2,2,3-Trimethylpentane
15.2,3,3-Trimethylpentane
Peaks
16.Toluene
17.Octane (C8)
18.Ethylbenzene
19.p-Xylene
20.2,3-Dimethylheptane
21.Nonane (C9)
22.5-Methylnonane
23.1,2-Methylethylbenzene
24.Decane (C10)
25.Undecane (C11)
26.1,2,3,5-Tetramethylbenzene
27.Naphthalene
28.n-Dodecane (C12)
29.1-Methylnaphthalene
30.Tridecane (C13)
image
GC_PC00774
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
using Rtx-5 DHA tuning column* 2-5 m, 0.25 mm ID,
SampleDHA/oxygenates setup blend
Injection
Inj. Vol.:0.1 µL split (split ratio 150:1)
Liner:4 mm ID deactivated cup inlet liner (cat.# 20835 [replaced by 20836])
Inj. Temp.:250 °C
Oven
Carrier GasH2, constant flow
Flow Rate:3.62 mL/min
Linear Velocity:55 cm/sec
DetectorFID @ 300 °C
Notes
*Note that the exact length of the tuning column was determined experimentally based on the resolution of critical pairs as per method D6730-01 (reapproved 2011).

Oven Temp.
A: 35 °C
B: 5 °C (hold 8.32 min) (elute C5) to 48 °C at 22 °C/min (hold 26.32 min) (elute ethylbenzene) to 141 °C at 3.20 °C/min (elute C12) to 300 °C at 1 °C/min


A: Front end of DHA/oxygenates setup blend
C5 Efficiency: 586,825 plates
C5 k': 0.476
tert-Butanol skew: 2.10
Resolution: tert-Butanol/2-methylbutene-2: 5.39


AcknowledgementChromatogram courtesy of Neil Johansen, Inc., Aztec, New Mexico, in association with Envantage Analytical Software, Inc., Cleveland, Ohio.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

DHA-100

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us