Simulated Distillation (Standard Calibration) on MXT-1 SimDist
Peaks | |
---|---|
1. | C6![]() |
2. | C7![]() |
3. | C8![]() |
4. | C9![]() |
5. | C10![]() |
6. | C11![]() |
7. | C12![]() |
8. | C14![]() |
Peaks | |
---|---|
9. | C16![]() |
10. | C18![]() |
11. | C20![]() |
12. | C24![]() |
13. | C28![]() |
14. | C32![]() |
15. | C36![]() |
16. | C40![]() |
17. | C44 |

GC_PC00224
Column | MXT-1 SimDist, 6 m, 0.53 mm ID, 0.15 µm (cat.# 70101) |
Sample | ASTM D2887 standard |
Conc.: | Wet needle |
Injection | cold on-column |
Temp. Program: | -17 °C to 433 °C |
Oven | |
Oven Temp.: | -12 °C to 430 °C at 20 °C/min |
Carrier Gas | H2, constant linear velocity |
Linear Velocity: | 40 cm/sec |
Detector | FID @ 430 °C |
Notes | FID sensitivity: 128 x 10-11 AFS |