Hydrocarbons (C7-C42) on Rtx-1
Peaks | |
---|---|
1. | C7![]() |
2. | C8![]() |
3. | C9![]() |
4. | C10![]() |
5. | C11![]() |
6. | C12![]() |
7. | C13![]() |
8. | C14![]() |
9. | C15![]() |
10. | C16![]() |
11. | C17![]() |
12. | C18![]() |
13. | C19![]() |
Peaks | |
---|---|
14. | C20![]() |
15. | C22![]() |
16. | C23![]() |
17. | C24![]() |
18. | C25![]() |
19. | C26![]() |
20. | C28![]() |
21. | C30![]() |
22. | C32![]() |
23. | C34![]() |
24. | C36![]() |
25. | C38![]() |
26. | C40![]() |
27. | C42 |

Column | Rtx-1, 30 m, 0.53 mm ID, 0.25 µm (cat.# 10125) |
Sample | Synthetic hydrocarbon mix |
Conc.: | ~0.1 mg/mL per component |
Injection | |
Inj. Vol.: | 0.2 µL Direct |
Liner: | Uniliner |
Inj. Temp.: | 340 °C |
Oven | |
Oven Temp.: | 40 °C to 340 °C at 5 °C/min |
Carrier Gas | H2, constant pressure |
Linear Velocity: | 40 cm/sec @ 40 °C |
Detector | FID @ 340 °C |
Notes | FID sensitivity: 64 x 10-11 AFS |