Comparison of Low-Level Pesticide Responses Using Restek Premium vs. Standard Deactivation Inlet Liners, Part I (QuEChERS Performance Standard C on Rxi-5ms)
Peaks | |
---|---|
1. | Vinclozoline![]() |
2. | Carbaryl![]() |
3. | Metalaxyl![]() |
4. | Methiocarb![]() |
5. | Cyprodinil![]() |
6. | Thiabendazole![]() |
7. | Captan![]() |
8. | Folpet![]() |

Column | Rxi-5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420) |
Sample | QuEChERS performance standard C (cat.# 31155) |
Diluent: | Toluene |
Conc.: | 100 pg/µL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Comparison of low-level pesticide responses using Premium single taper with wool (cat.# 23303.5) and standard deactivation single taper with wool (cat.# 22406) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 330 °C at 21 °C/min (hold 1.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.5 mL/min |
Detector | FID @ 340 °C |
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |