Innovative Chromatography Products
www.restek.it
www.restek.it

Comparison of Low-Level Pesticide Responses Using Restek Premium vs. Standard Deactivation Inlet Liners, Part I (QuEChERS Performance Standard C on Rxi-5ms)
Peaks | |
---|---|
1. | Vinclozoline![]() |
2. | Carbaryl![]() |
3. | Metalaxyl![]() |
4. | Methiocarb![]() |
5. | Cyprodinil![]() |
6. | Thiabendazole![]() |
7. | Captan![]() |
8. | Folpet![]() |

GC_GN1185
Column | Rxi-5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420) |
Sample | QuEChERS performance standard C (cat.# 31155) |
Diluent: | Toluene |
Conc.: | 100 pg/µL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Comparison of low-level pesticide responses using Premium single taper with wool (cat.# 23303.5) and standard deactivation single taper with wool (cat.# 22406) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 330 °C at 21 °C/min (hold 1.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.5 mL/min |
Detector | FID @ 340 °C |
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.