Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

GC Multiresidue Pesticide Standard #6-SPP on Rxi-5ms by GC-MS

PeakstR (min)
1.Tefluthrin14.23
2.Transfluthrin15.18
3.Anthraquinone16.02
4.Bioallethrin17.17
5.Resmethrin 1*20.43
6.Resmethrin 2*20.55
7.Tetramethrin 1*21.00
8.Tetramethrin 2*21.14
9.Bifenthrin21.15
10.Phenothrin 1*21.59
11.Phenothrin 2*21.71
12.lambda-Cyhalothrin22.30
13.Acrinathrin22.51
14.cis-Permethrin23.14
15.trans-Permethrin23.29
PeakstR (min)
16.Cyfluthrin 1*23.83
17.Cyfluthrin 2*23.93
18.Cyfluthrin 3*24.02
19.Cyfluthrin 4*24.06
20.Cypermethrin 1*24.19
21.Cypermethrin 2*24.30
22.Cypermethrin 3*24.39
23.Cypermethrin 4*24.43
24.Flucythrinate 1*24.43
25.Flucythrinate 2*24.66
26.Fenvalerate 1*25.25
27.tau-Fluvalinate 1*25.47
28.Fenvalerate 2*25.48
29.tau-Fluvalinate 2*25.53
30.Deltamethrin26.09
*Isomers numbered according to elution order.
image
GC_FS0605
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleGC multiresidue pesticide standard #6-SPP (cat.# 32568)
Diluent:Toluene
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1555-5507
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Electron Energy:70 eV
Solvent Delay Time:5 min
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesBioallethrin isomers are only slightly resolved with this method, so they are treated as one peak.

Reconstructed chromatogram from select ions.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.