trans-Fats in Frosting on Rt-2560

PeakstR (min)
1.C18:025.032
2.C18:1 (trans-6)26.576
3.C18:1 (trans-9)26.702
4.C18:126.835
5.C18:1 (trans-11)26.937
6.C18:1 (cis-6)27.201
7.C18:1 (cis-9)27.372
8.C18:1 27.529
9.C18:1 (cis-11)27.751
10.C18:128.045
11.C18:2 (cis-9,12)31.011
image
GC_FS0567
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleSee notes
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:160 °C (hold 20 min) to 250 °C at 2 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesSample: 0.12 g of chocolate frosting was dissolved in 1 mL hexane and 0.5 mL acetone and then reacted with 75 µL NaOCH3 (2 M). The sample was then vortexed for 10 sec, and the reaction was stopped by the addition of 1 mL of acetic acid (0.5 M). FAMEs were extracted into the hexane layer, which was injected for analysis.

Standard: cis/trans FAME mix (cat.# 35079) was diluted 50x in hexane, resulting in a final FAME concentration of 0.16–0.40%.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

partially hydrogenated oils

,

FAMEs

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.