Food Industry FAMEs on Rt-2560 by AOCS Method Ce-1j-07 Using H2 (Optimized Method)
Peaks
tR (min)
Conc. (µg/mL)
Structural Nomenclature
1.
Methyl caproate
4.740
400
C6:0
2.
Methyl octanoate
4.953
400
C8:0
3.
Methyl decanoate
5.317
400
C10:0
4.
Methyl undecanoate
5.587
200
C11:0
5.
Methyl dodecanoate
5.939
400
C12:0
6.
Methyl tridecanoate
6.398
200
C13:0
7.
Methyl myristate
6.998
400
C14:0
8.
Methyl myristoleate
7.736
200
C14:1 (c9)
9.
Methyl pentadecanoate
7.781
200
C15:0
10.
Methyl pentadecenoate
8.734
200
C15:1 (C10)
11.
Methyl palmitate
8.806
600
C16:0
12.
Methyl palmitoleate
9.831
200
C16:1 (c9)
13.
Methyl heptadecanoate
10.136
200
C17:0
14.
Methyl heptadecenoate
11.448
200
C17:1 (c10)
15.
Methyl stearate
11.882
400
C18:0
16.
Methyl octadecenoate
12.877
200
C18:1 (t9)
17.
Methyl oleate
13.345
400
C18:1 (c9)
18.
Methyl linolelaidate
14.685
200
C18:2 (t9,t12)
19.
Methyl linoleate
15.928
200
C18:2 (c9,c12)
20.
Methyl arachidate
17.136
400
C20:0
21.
Methyl linolenate
18.252
200
C18:3 (c6,c9,c12)
22.
Methyl eicosenoate
19.422
200
C20:1 (c11)
23.
Methyl linolenate
19.740
200
C18:3 (c9,c12,c15)
24.
Methyl heneicosanoate
20.592
200
C21:0
25.
Methyl eicosadienoate
21.621
200
C20:2 (c11,c14)
26.
Methyl behenate
22.380
400
C22:0
27.
Methyl eicosatrienoate
22.884
200
C20:3 (c8,c11,c14)
28.
Methyl erucate
23.510
200
C22:1 (c13)
29.
Methyl eicosatrienoate
23.656
200
C20:3 (c11,c14,c17)
30.
Methyl arachidonate
23.901
200
C20:4 (c5,c8,c11,c14)
31.
Methyl tricosanoate
24.240
200
C23:0
32.
Methyl docosadienoate
25.408
200
C22:2 (c13,c16)
33.
Methyl eicosapentaenoate
26.293
200
C20:5 (c5,c8,c11,c14,c17)
34.
Methyl lignocerate
26.293
400
C24:0
35.
Methyl nervonate
27.653
200
C24:1 (c15)
36.
Methyl docosahexaenoate
33.053
200
C22:6 (c4,c7,c10,c13,c16,c19)
GC_FF1348
Column
Rt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
Sample
Food industry FAME mix (cat.# 35077)
Diluent:
Hexane
Conc.:
10,000 µg/mL total concentration
Injection
Inj. Vol.:
1 µL split (split ratio 20:1)
Liner:
Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:
250 °C
Oven
Oven Temp.:
180 °C (hold 19.8 min) to 215 °C at 32.3 °C/min (hold 11 min)
Carrier Gas
H2, constant flow
Flow Rate:
2.5 mL/min
Detector
FID @ 250 °C
Constant Column + Constant Make-up:
52 mL/min
Hydrogen flow:
40 mL/min
Air flow:
400 mL/min
Data Rate:
50 Hz
Instrument
Agilent 7890A GC
Notes
Method optimized from He using the EZGC method translator (https://blog.restek.com/?p=52224). C4:0 Methyl butyrate (623-42-7) elutes in the solvent front. Resolution of critical pair cis-11-C20:1 (#22) and cis-9,12,15-C18:3 (#23) is 2.03.
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.
I cookies consentono varie funzionalità che renderanno più agevole la tua visita al sito Restek. Utilizzando questo sito, acconsenti all’uso dei cookies in accordo con le nostre linee guida. Per saperne di più, vedi la nostra politica sulla privacy.
Restek Domestic Customer Service
Grazie.
Il messaggio è stato inviato. La ricontatteremo non appena possibile.
Messaggio non inviato.
Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037