Food Industry FAMEs on Rt-2560 by AOCS Method Ce-1j-07 using H2 (Original Method)

PeakstR (min)Conc. (µg/mL)Structural Nomenclature
1.Methyl caproate4.741400C6:0
2.Methyl octanoate4.954400C8:0
3.Methyl decanoate5.319400C10:0
4.Methyl undecanoate5.589200C11:0
5.Methyl dodecanoate5.942400C12:0
6.Methyl tridecanoate6.403200C13:0
7.Methyl myristate7.005400C14:0
8.Methyl myristoleate7.744200C14:1 (c9)
9.Methyl pentadecanoate7.790200C15:0
10.Methyl pentadecenoate8.745200C15:1 (C10)
11.Methyl palmitate8.818600C16:0
12.Methyl palmitoleate9.846200C16:1 (c9)
13.Methyl heptadecanoate10.153200C17:0
14.Methyl heptadecenoate11.469200C17:1 (c10)
15.Methyl stearate11.905400C18:0
16.Methyl octadecenoate12.903200C18:1 (t9)
17.Methyl oleate13.372400C18:1 (c9)
18.Methyl linolelaidate14.718200C18:2 (t9,t12)
19.Methyl linoleate15.963200C18:2 (c9,c12)
20.Methyl arachidate17.176400C20:0
21.Methyl linolenate18.295200C18:3 (c6,c9,c12)
22.Methyl eicosenoate19.470200C20:1 (c11)
23.Methyl linolenate19.787200C18:3 (c9,c12,c15)
24.Methyl heneicosanoate21.060200C21:0
25.Methyl eicosadienoate23.742200C20:2 (c11,c14)
26.Methyl behenate26.204400C22:0
27.Methyl eicosatrienoate27.448200C20:3 (c8,c11,c14)
28.Methyl erucate29.836200C22:1 (c13)
29.Methyl eicosatrienoate29.996200C20:3 (c11,c14,c17)
30.Methyl arachidonate30.667200C20:4 (c5,c8,c11,c14)
31.Methyl tricosanoate32.652200C23:0
32.Methyl docosadienoate34.142200C22:2 (c13,c16)
33.Methyl eicosapentaenoate34.934200C20:5 (c5,c8,c11,c14,c17)
34.Methyl lignocerate35.417400C24:0
35.Methyl nervonate36.878200C24:1 (c15)
36.Methyl docosahexaenoate42.513200C22:6 (c4,c7,c10,c13,c16,c19)
image
GC_FF1347
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleFood industry FAME mix (cat.# 35077)
Diluent:Hexane
Conc.:10,000 µg/mL total concentration
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:180 °C (hold 32 min) to 215 °C at 20 °C/min (hold 30.25 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesHydrogen flow optimized to achieve effective linear velocity (https://blog.restek.com/?p=52224).
C4:0 Methyl butyrate (623-42-7) elutes in the solvent front.
Resolution of critical pair cis-11-C20:1 (#22) and cis-9,12,15-C18:3 (#23) is 2.03.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Oil

,

Fat

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us