Extra Virgin Olive Oil on Rxi-65TG (30 m x 0.32 mm x 0.1 µm)

PeakstR (min)
1.1,3-Palmitin-2-olein (POP)14.592
2.1,2-Palmitin-3-linolein (PPL)14.875
3.1-Palmitin-2-olein-3-stearin (POS)16.324
4.1,2-Olein-3-palmitin (POO)16.649
PeakstR (min)
5.1-Palmitin-2-linolein-3-olein (PLO)17.013
6.1,2-Linolein-3-palmitin (PLL)17.397
7.1,2-Stearin-3-olein (SOS)18.348
8.Triolein (OOO)19.065
image
GC_FF1343
ColumnRxi-65TG, 30 m, 0.32 mm ID, 0.10 µm (cat.# 17109)
SampleExtra virgin olive oil
Diluent:Isooctane
Conc.: 5 mg/mL
Injection
Inj. Vol.:1 µL split (split ratio 25:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:360 °C
Oven
Oven Temp.:200 °C (hold 1.2 min) to 350 °C at 21.7 °C/min to 365 °C at 1.2 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:1.28 mL/min
Dead Time:1.4331 min
DetectorFID @ 365 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC
NotesSample Preparation: 50 mg of extra virgin olive oil was diluted to 10 mL with isooctane.
Download PDF