Tea Tree Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

PeakstR (min)
1.α-Pinene0.514
2.Terpinolene0.612
3.p-Cymene0.622
4.D-Limonene0.625
5.Eucalyptol0.640
6.γ-Terpinene0.674
7.Pinocarveol0.840
8.Terpinen-4-ol0.926
PeakstR (min)
9.α-Terpineol0.945
10.Nerol0.991
11.Geraniol1.043
12.γ-Elemene1.274
13.Caryophyllene1.513
14.Aromadendrene1.562
15.α-Caryophyllene1.604
16.Alloaromadendrene1.617
17.Ledene1.691
image
GC_FF1339
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleTea tree oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).