Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Peppermint Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)

PeakstR (min)
1.α-Pinene1.253
2.3-Methylcyclohexanone1.304
3.β-Phellandrene1.367
4.β-Pinene1.399
5.p-Cymene1.551
6.D-Limonene1.570
7.Eucalyptol1.599
8.Methone 12.231
PeakstR (min)
9.Methone 22.274
10.Neoisomenthol2.295
11.Menthol2.364
12.Isomenthol2.417
13.α-Terpineol2.440
14.Pulegone2.706
15.Menthyl acetate3.006
16.Caryophyllene3.953
image
GC_FF1326
ColumnRxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602)
SamplePeppermint oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Essential oils

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.