Patchouli Oil on Rxi-5Sil MS

PeakstR (min)
1.α-Pinene2.062
2.β-Pinene2.290
3.D-Limonene2.545
4.δ-Elemene5.246
5.β-Elemene5.829
6.β-Patchoulene5.867
7.Caryophyllene6.222
8.α-Guaiene6.344
PeakstR (min)
9.Seychellene6.596
10.α-Patchoulene6.710
11.β-Guaiene6.741
12.Patchoulene6.795
13.Cedrene6.991
14.δ-Guaiene7.069
15.Valencene8.691
16.Patchoulol8.937
image
GC_FF1312
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SamplePatchouli oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 11 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.31 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Essential oils

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.