Lemongrass Oil on Rtx-Wax

PeakstR (min)
1.Camphene1.543
2.D-Limonene1.806
3.β-Ocimene1.858
4.α-Ocimene1.916
5.4-Nonanone2.180
6.Methyl heptenone2.212
PeakstR (min)
7.Linalool3.179
8.Caryophyllene3.636
9.Z-Citral4.038
10.E-Citral4.340
11.Geranyl acetate4.490
12.α-Amorphene4.588
13.Geraniol4.997
image
GC_FF1306
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
SampleLemongrass oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:230 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

essential oils

,

citral

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us