Citrals on Rtx-Wax
Peaks | Rtx-Wax tR | Stabilwax tR | |
---|---|---|---|
1. | Citral B![]() | 18.30 | 18.53 |
2. | Citral A![]() | 18.99 | 19.22 |

GC_FF1280
Column | See notes |
Sample | Custom citral standard |
Diluent: | Acetone |
Conc.: | 1000 ppm |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 106 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 250 °C at 8 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.1 mL/min |
Linear Velocity: | 19 cm/sec @ 40 °C |
Detector | FID @ 250 °C |
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Column Details Rtx-Wax, 30 m, 0.32 mm, 0.50 μm (cat.# 12439) Stabilwax, 30 m, 0.32 mm, 0.50 μm (cat.# 10639) |