Rum on Rxi-1301Sil MS

PeakstR (min)
1.Acetaldehyde4.370
2.Methanol4.474
3.Ethanol5.395
4.n-Propanol6.836
PeakstR (min)
5.Ethyl acetate7.403
6.Isobutanol8.132
7.Acetal9.458
8.3-Methyl-1-butanol (isoamyl alcohol)10.308
9.2-Methyl-1-butanol (active amyl alcohol)10.379
image
GC_FF1279
ColumnRxi-1301Sil MS, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16097)
SampleRum
Conc.:Neat
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 300 °C at 10 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1420-3007.3
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesWater coelutes with acetaldehyde and methanol, which results in distortion of the baseline.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

fusel alcohol

,

liquor analysis

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.