Electronic Cigarette Liquid on Rtx-VMS (GC-MS)

Column | Rtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915) | ||||||||
Sample | |||||||||
Diluent: | Methylene chloride | ||||||||
Conc.: | Electronic cigarette liquid diluted 2:1 | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 10:1) | ||||||||
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) | ||||||||
Inj. Temp.: | 250 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 35 °C (hold 1 min) to 250 °C at 11 °C/min (hold 4 min) | ||||||||
Carrier Gas | He, constant flow | ||||||||
Flow Rate: | 2.0 mL/min | ||||||||
Linear Velocity: | 51.15 cm/sec @ 35 °C | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 250 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Extractor | ||||||||
Extractor Lens: | 6mm ID | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Tune Type: | BFB | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890B GC & 5977A MSD |